F620 : FEATURES

  • Differential pressure decay leak measurement
  • Large range of leak measurement (∆P) F.S.: 20Pa, 50 Pa, 500 Pa or 5000 Pa
  • Mechanical or electronic regulator from vacuum to 20 bar
  • 128 programs
  • Setting on line (via USB)
  • 2 languages (English + other one)
  • Manual calibration on front panel with calibrated leak
  • Data storage on internal memory or USB key

TEST MODES :

  • ∆P (Pa, 1/10 Pa)
  • ∆P/t (Pa/s, 1/10 Pa/s)
  • Flow units (sccm, mm3/s, cm3/s, cm3/min, cm3/h…)
  • Blockage test
  • Burst test
  • Sealed components test and others, depending on your applications

ENVIRONMENT :

ROHS standard

FLEXIBLE COMMUNICATION FORMAT :

  • RS232: printer, modbus RTU
  • Slave USB: PC supervision (WinAteq 300, sesame)
  • Master USB: USB key to transfert (parameters, results, …)
  • Option Fieldbus: profibus, devicenet, profinet, Ethernet/IP

MEASUREMENT CHARACTERISTICS :

PRESSURE DROP MEASUREMENT
Range Accuracy up to* Max. Resolution
20Pa** ± (2.0 % Reading + 0.05 Pa) 0.0001 Pa
50Pa  ± (1.0 % Reading + 0.5 Pa) 0.01 Pa
500Pa  ± (0.5 % Reading + 0.5 Pa) 0.1 Pa
5000Pa  ± (0.5 % Reading + 5 Pa) 1 Pa

Accuracy: linearity + repeatability + hysteresis *Optional laboratory accuracy **Maximum Test Pressure 3 bar