ATEQ CDF60 portable flow calibrator front panel
Product highlights: pocket size and user-friendly display, Li-ion rechargeable battery, USB interface, high accuracy

CDF60

HAND-HELD LEAK/FLOW CALIBRATOR

The CDF60 calibrator checks the calibration of leak and flow ATEQ instruments as well as any other instrument.

This light, compact and user friendly calibrator is essential for field or laboratory checks, when accuracy and repeatability cannot be compromised and large instrument are too bulky or too expensive.

With the CDF60, you can now easily check and follow the requirements of ISO or other quality standards.

CDF60 : APPLICATIONS

1. ATEQ LEAK/FLOW INSTRUMENTS

Calibration check of all ATEQ Leak Testers

2. COMPETITOR LEAK/FLOW TESTERS

Calibration check on any type of Leak Testers

ATEQ CDF60 flow calibrator kit with handheld meter, red carrying case, USB cable, charger, and user manual

CDF60 : FEATURES

  • Atmospheric pressure and temperature indicator
  • Automatic sleep mode
  • Print out (USB or RS232)
  • battery level check
  • LCD display (4 x 20 characters)
  • 1 year warranty
  • Accessories : Carrying case, needle valve, soft case, protective holster…
ATEQ CDF60 handheld calibrator alongside an ATEQ F620 leak tester
FLOW MEASUREMENT
Range Accuracy up to* Max. Resolution
4 ml/min ± 2.0 % Reading + 0.01 ml/min

(± 1.0 % Reading + 0.01 ml/min)

0.001 ml/min
40 ml/min  ± 2.0 % Reading + 0.1 ml/min

(± 1.0 % Reading + 0.1 ml/min)*

0.01 ml/min

Rdg : Reading Value, Accuracy in ambient conditions, defined as non-linearity, hysteresis, repeatability, temperature, effect and long-term drift.

*Specification in option (calibrated in our ISO/IEC 17025 laboratory

CDF60 : SPECIFICATIONS

  • TEMPERATURE

    • Operating : + 10°C at + 45°C
    • Storage : 0°C at + 60°C
  • AIR SUPPLY

    • Clean and dry air
    • Air quality standard to be applied (ISO 8573-1)